August 30, 2010
Re: Addendum to STA® Compact, STA® Compact CT, STA-R® and STA-R® Evolution Using Soft Stylets: Preventative Maintenance Optimization and Operator's Security
Dear Valued Customer,
Please find, in the link above, an addendum that details a change in maintenance for your instrument. Stago continues to evaluate and improve our processes and has found that the continued use of metal stylets to clean out the needles on a weekly basis is not required.
As per the addendum, only use the soft stylet (Item number 80343) if you suspect a clog. Do not pass the stylet through a cap-piercing needle.
MKT-10-022, Eff. Date 08/30/2010