Download MKT-10-022.doc

Download Soft_Stylet_mandrell_CR-209-052-Attachment02.pdf

Stago

Contact:
Alicia Carpenter, MT(ASCP)
Technical Training Manager
1-800-222-2624 x4198



August 30, 2010

Re: Addendum to STA® Compact, STA® Compact CT, STA-R® and STA-R® Evolution Using Soft Stylets: Preventative Maintenance Optimization and Operator's Security


Dear Valued Customer,


Please find, in the link above, an addendum that details a change in maintenance for your instrument. Stago continues to evaluate and improve our processes and has found that the continued use of metal stylets to clean out the needles on a weekly basis is not required.

As per the addendum, only use the soft stylet (Item number 80343) if you suspect a clog. Do not pass the stylet through a cap-piercing needle.



MKT-10-022, Eff. Date 08/30/2010